The unknown samples S1, and S2 were analysed on Gas Chromatograph (GC) equipped with Electron Capture Detector (ECD) and Nitrogen Phosphorous Detector (NPD) using the capillary column. The sample S1 was found to contain -endosulfan (14.8 + 02.3 mg ) and -endosulfan (17.0 + 13.2 mg). S2 was found to contain primicarb (13.8 + 07.8 mg) and azin phosmethyl (32.6 + 03.5 mg). The result of all four pesticides were found to be satisfactory except primicartb, hanging little bit for residue analysis, when compared with actual amount i.e. -endosulfan (11.0 mg), -endosulfan (16.9 mg), primicarb (24.0 mg) and azin phosmethyl (32.0 mg) furnished by the Danish Institute of Plant and soil science, Denmark.

Tahir Anwar, Seema Tahir, Imtiaz Ahmad. (2006) INTERLABORATORY COMPARISON OF UNKNOWN SAMPLES OF INSECTICIDES ON GAS CHROMATOGRAPH USING CAPILLARY COLUMNS, , Volume 3, Issue 4.